eVHP High Pressure Vessel Set…
eVHP High Pressure Vessel Set
Descripción
• Total 18 eVHP vessels.
• Min recommended vessels used – 04.
• Each Vessel volume 100mL.
• Recommended sample quantity 12-50mL, depending on sample.
• Design temperature 300°C.
• Maximum operating temperature 260°C.
• Normal operating temperature 230°C.
• Design pressure ~165 bar/ 2400 psi.
• Maximum pressure ~83 bar/ 1200 psi.
• Operating pressure control at ~42 bar/ 610 psi (Op Guard limit).
• Includes vessel carousel.
• Includes vessel loading station.
• Every vessel is monitored individually for temperature and pressure.
• Vessel over-pressure management system has controlled release of over pressure for full operational temperature range. Each vessel is individually tested for 600 psi (40 bar) pressure.
Equipped with resealing OPGuard™ vessel protection system.